The Dragon X is designed to carry the X-ray unit from door to door of patient room in clinics and hospitals and therefore, can be reduced patient positioning time and efforts.
Standard applications are: medical radiographic use in nursing homes, private homes, field clinics, correctional facilities, military field operations, intensive care units, emergency units, extremity orthopaedic or anywhere an x-ray machine must be brought to a patient.
The system can be ordered with two different type of console: Touch screen option with APR 256 programs or Overlay option also with APR and 20 configurable programs.
RADIOGRAPHIC | GENERATOR MODEL SPSL8HC |
---|---|
KW Maximum Power | 8 KW Single-Phase 220/240VAC 50/60 HZ |
kVp Range ( 1kVp steps) | 40 – 125 Kvp Accuracy: ±(3% + 1 kVp) |
mA Range | 5 to 100mA in 16 steps (Renard scale) 5, 6,4, 8, 10, 12,5, 16, 20, 25, 32, 40, 50, 64, 80, 100 mA Accuracy: ±(4% + 1 mA) |
Exposure Time Range | 0.001 – 10 seconds steps according to R´10 |
Accuracy: ± (2% + 0.1 ms) | |
mAs Range | 0.1 mAs – 250 mAs steps according to R´10 Accuracy: ± (5% + 0.1 mAs) |